Title |
Thiocyanate determination by ion pair chromatography / |
Translation of Title |
Tiocianato nustatymas jonų porų chromatografijos metodu. |
Authors |
Tamošiūnas, Vytautas ; Padarauskas, Audrius ; Pranaitytė, Birutė |
Full Text |
|
Is Part of |
Chemija.. Vilnius : Lietuvos mokslų akademijos l-kla. 2006, t. 17, Nr. 2-3, p. 21-24.. ISSN 0235-7216 |
Keywords [eng] |
ion pair chromatography ; thiocyanate ; human saliva |
Abstract [eng] |
A simple and rapid ion pair chromatographic method has been developed for the determination of thiocyanate. Separation was carried out on a 25 cm × 4.6 mm Ascentis C18 column (5 μm particle size) with a mobile phase containing 2 mmol/l tetrabutylammonium (TBA)-phosphate (pH 6.5) in 40 (v/v) % acetonitrile at a flow rate of 0.5 ml/min. Detection was performed by direct UV absorption at 210 nm. This study investigated the effect of various experimental factors, such as acetonitrile content, pH, and ion-pairing cation concentration in the mobile phase on the retention behaviour of anions. Under optimized conditions, four anions (Br-, NO3 - , I- and SCN-) were effectively separated within 8 minutes. The calibration curve was linear in the thiocyanate concentration range from 5×10-6 to 5×10-4 mol/l with an excellent correlation coefficient of more than 0.999. The limit of detection was 1.5×10-6 mol/l (signal / noise = 3). The method was applied for determination of thiocyanate ions in human saliva. Saliva sample preparation involved only centrifugation and appropriate dilution. |
Published |
Vilnius : Lietuvos mokslų akademijos l-kla |
Type |
Journal article |
Language |
English |
Publication date |
2006 |